<ErrorNumber>21001</ErrorNumber>
<ErrorFamily>SCSI</ErrorFamily>
<ErrorMessage>
SCSI: SCSI parity error while imaging, check plate for artifacts
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>All</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The firmware checks the parity of each word of image data sent by the workstation. If there is a
discrepancy, this unsolicited error is reported at the end of the image. Note that a parity error does not
cause the image to abort or suspend, so plates should be inspected when this message is reported.
</text></DescriptionText>
<DescriptionText><text>
The SCSI hardware should be checked as described in the Service Manual's Troubleshooting section. If
the SCSI hardware does not seem to be the problem, the possibility of external electrical noise corrupting
the SCSI data should be investigated.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
The plate should be inspected for artifacts or it should be re-imaged if there is concern that any artifacts
would be difficult to detect on the plate.
</text></DescriptionText>
</OperatorAction>
<Cause>
Faulty SCSI hardware
</Cause>
<Cause>
External electrical noise
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:50 2008 PST
</Updated>
<Links>
</Links>
</Error>